Fit Data
Choose the type of Curve Fit to be performed on the Calibration data. There are
4 options available.
Linear Fit Process Data
Curve Fit Process Data
Linear Fit Factory Data
Curve Fir Factory Data
Copy Factory Table
The selected sensor's factory calibration table is copied to process calibration
table.
6.6.2.2
Factory Calibration
Build Calibration
View/Edit Calibration
Build Calibration
For each Sensor, up to 8 different samples can be used during the calibration
process.
Select the Sensor number by choosing the
Sensor
Toggle Button.
Select the Sample number by choosing on the
Sample
Toggle Button.
Enter the
User Value
(Ex: - Known concentration value of a Standard) in the text
box provided. This value can also be changed using
View/Edit Calibration
menu.
Choose
Read New Data
to capture the data.
Choose
Store New Value
to store the data.
Repeat steps 2 to 5 for each sample.