-73-
Test Relays
This Menu displays the current status of the FOUR Relays i.e. Signal to which
Relay has been assigned and whether it is switched ON / OFF.
Choose the ON / OFF Toggle button to Switch ON / OFF the Relay.
6.6.5.5
Clear/Reset Menu
System Reset
Default Values
Clear Process Calibration
Clear Factory Calibration
Clear Results
Reset Alarms
Choose
System Reset
to reset the Hardware.
Choose
Default Values to
default the parameters of the Messenger.
On Selection of Default Values, Factory calibration table will not be defaulted.
Clear Process Calibration
Choose
Sensor (1...4).
On Selection of Sensor, choose
Copy From Factory Table
key. Factory
calibration table contents of the Selected Sensor are copied to the Process
calibration table.
On Selection of Sensor, choose
Clear Table
key. Process calibration table
contents of Selected Sensor are defaulted to 0.
Clear Factory Calibration
Choose
Sensor (1...4)
On Selection of Sensor when
Clear Table
key is entered. Factory calibration
table contents of selected sensor are defaulted to 0.