Chapter 5: Setup
Fresenius Medical Care 5008 SM 1/08.04
5-7
O
Auto On
Requirements:
– Pre-program the T1 test under Auto On.
(Observe the time programming of the osmosis installation.)
– CDS for acid connected.
Submenu
Default value
Value range
Resolution
Selectable
options
Weekly programs
The program and the power-up time may be preselected.
Then turn programming on or off via Status.
If various programming actions have been performed, it is possible to turn them all
on or off via the
Auto On Programs I/O
button.
Program
No program
–
–
Rinse
Heat disinfection
T1 Test
No program
Power-up time
00:00
00:00–24:00
1 min
–
Single programs
The program and the power-up time may be preselected.
Then turn programming on or off via Status.
If various programming actions have been performed, it is possible to turn them all
on or off via the
Auto On Programs I/O
button.
Program
No program
–
–
Rinse
Heat disinfection
T1 Test
No program
Power-up time
00:00
00:00–24:00
1 min
–
The bi
b
ag
®
may be installed after completion of the last
disinfection of the 5008 hemodialysis system (72 hours
maximum before the treatment).
For profiting from this possibility, observe the following notes.
Caution
After removal of the foil, immediately connect the bi
b
ag
®
using aseptic
techniques. Then close the bicarbonate flap.
Summary of Contents for 5008
Page 2: ......
Page 6: ...Kapitel 1 Index 1 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 10: ...Chapter 2 Important Information 2 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 22: ...Chapter 3 Specifications 3 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 52: ...Chapter 5 Setup 5 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 60: ...Chapter 5 Setup 5 20 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 66: ...Chapter 6 TSC TMC Maintenance 6 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 74: ...Chapter 6 TSC TMC Maintenance 6 10 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 90: ...Chapter 6 TSC TMC Maintenance 6 26 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 96: ...Chapter 7 Error Messages 7 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 107: ...Chapter 10 Repair Fresenius Medical Care 5008 SM 1 08 04 10 1 10 Repair no contents yet ...
Page 108: ...Chapter 10 Repair 10 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 110: ...Chapter 11 Functional Description 11 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 128: ...Chapter 12 Service Program Option 12 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 130: ...Chapter 12 Service Program Option 12 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 132: ...Chapter 12 Service Program Option 12 6 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 150: ...Chapter 12 Service Program Option 12 24 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 164: ...Chapter 12 Service Program Option 12 38 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 166: ...Chapter 12 Service Program Option 12 40 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 168: ...Chapter 12 Service Program Option 12 42 Fresenius Medical Care 5008 SM 1 08 04 ...