Chapter 4: Installation
Fresenius Medical Care 5008 SM 1/08.04
4-11
4
Check – dialysate flow
4.1
In the service mode, select CALIBRATE.
Basic requirements:
The hemodialysis system
must be closed.
Flow on, flow 800 ml/min
Check flow.
Desired value: 770 ml/min to 830 ml/min
5
Check – temperature
5.1
In the service mode, select CALIBRATE.
Basic requirements:
The hemodialysis system
must be closed.
Temperature 37 °C, flow on, flow 500 ml/min,
Response time approx. 10 min.
Check PT07 (temperature).
Desired value: 36.8 °C to 37.2 °C (display on hemodialysis system)
Measure the reference temperature with an external measuring instrument.
Difference = Reference temperature minus PT07
Desired value – difference: –0.5 °C to +0.2 °C
Example:
PT07: 37 °C
Desired value reference temperature: 36.5 °C to 37.2 °C
6
Check – conductivity
6.1
In the service mode, select CALIBRATE.
Basic requirements:
The hemodialysis system
must be closed.
External measuring instrument (e.g. UMED) connected for at least 5 minutes.
Temperature 37 °C, flow on
Check CD7 (conductivity).
Desired value: approx. 13.5 mS/cm to approx. 14.5 mS/cm
Measure the reference conductivity with an external measuring instrument.
Difference = Reference conductivity minus CD7
Desired value – difference:
±
0.2 mS/cm
7
Check – blood leak detector
In the service mode, select CALIBRATE.
Basic requirements:
The hemodialysis system
must be closed. (Avoid external light.)
Temperature of approx. 37 °C achieved, flow on, flow 500 ml/min,
7.1
Check the blood leak:
Desired value: 4.8 V to 5.2 V
7.3
Check the dimness:
Desired value: 4.8 V to 5.2 V
No.
Description
Summary of Contents for 5008
Page 2: ......
Page 6: ...Kapitel 1 Index 1 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 10: ...Chapter 2 Important Information 2 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 22: ...Chapter 3 Specifications 3 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 52: ...Chapter 5 Setup 5 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 60: ...Chapter 5 Setup 5 20 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 66: ...Chapter 6 TSC TMC Maintenance 6 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 74: ...Chapter 6 TSC TMC Maintenance 6 10 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 90: ...Chapter 6 TSC TMC Maintenance 6 26 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 96: ...Chapter 7 Error Messages 7 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 107: ...Chapter 10 Repair Fresenius Medical Care 5008 SM 1 08 04 10 1 10 Repair no contents yet ...
Page 108: ...Chapter 10 Repair 10 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 110: ...Chapter 11 Functional Description 11 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 128: ...Chapter 12 Service Program Option 12 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 130: ...Chapter 12 Service Program Option 12 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 132: ...Chapter 12 Service Program Option 12 6 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 150: ...Chapter 12 Service Program Option 12 24 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 164: ...Chapter 12 Service Program Option 12 38 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 166: ...Chapter 12 Service Program Option 12 40 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 168: ...Chapter 12 Service Program Option 12 42 Fresenius Medical Care 5008 SM 1 08 04 ...