Chapter 3: Specifications
3-4
Fresenius Medical Care 5008 SM 1/08.04
Service/diagnostics
(Protected by cover!)
For inhouse computer diagnostics.
Port: DSUB 15-pin
24 V
(Protected by cover!)
24 V connection (2 A fuse)
Port: Flanged socket, 4-pin
Alarm output
For the connection of an external alarm indicator (nurse call). (Potential-
free alarm output. Alternating contact maximum 24 V/24 W).
Port: 5-pin diode plug via a shielded line; shield grounded on either side.
O
Override conditions
When overriding a safety system the responsibility for the patient’s
safety rests with the operator of the machine.
Audible alarm suppression
Mute alarm time: maximum 2 minutes
(adjustable in the SETUP)
Alarm override
After confirmation of the error message and start of the blood systems:
Arterial and venous pressure alarm for approx. 10 seconds (window
inactive)
Air detector alarm for approx. 2 seconds
Blood leak override
Override time: maximum 2 minutes
Override air-bubble
detector
Override time: after starting removal of air: approx. 4 seconds
O
Operating programs
T1 test
Automatic test for verification of the operating and safety systems.
The T1 test is mandatory,
– after power on (not following a power failure)
– after a cleaning program
Preparation
Defined by the optical detector located below the venous bubble
catcher.
Preparation is terminated as soon as the optical detector senses
opaque fluid in the blood lines.
Priming and rinsing the
blood lines
Minimum rinse volume 500 ml; automatic switching to rinsing, if level in
bubble catcher detected. Automatic raising of the fluid level during the
rinse phase.
Reinfusion
Reinfusion volume adjustable in the SETUP.
Return to dialysis still possible.
Dialysis
Bicarbonate dialysis
ISO-UF
Ultrafiltration without dialysate flow (Bergström method)
Summary of Contents for 5008
Page 2: ......
Page 6: ...Kapitel 1 Index 1 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 10: ...Chapter 2 Important Information 2 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 22: ...Chapter 3 Specifications 3 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 52: ...Chapter 5 Setup 5 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 60: ...Chapter 5 Setup 5 20 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 66: ...Chapter 6 TSC TMC Maintenance 6 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 74: ...Chapter 6 TSC TMC Maintenance 6 10 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 90: ...Chapter 6 TSC TMC Maintenance 6 26 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 96: ...Chapter 7 Error Messages 7 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 107: ...Chapter 10 Repair Fresenius Medical Care 5008 SM 1 08 04 10 1 10 Repair no contents yet ...
Page 108: ...Chapter 10 Repair 10 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 110: ...Chapter 11 Functional Description 11 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 128: ...Chapter 12 Service Program Option 12 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 130: ...Chapter 12 Service Program Option 12 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 132: ...Chapter 12 Service Program Option 12 6 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 150: ...Chapter 12 Service Program Option 12 24 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 164: ...Chapter 12 Service Program Option 12 38 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 166: ...Chapter 12 Service Program Option 12 40 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 168: ...Chapter 12 Service Program Option 12 42 Fresenius Medical Care 5008 SM 1 08 04 ...