Chapter 3: Specifications
Fresenius Medical Care 5008 SM 1/08.04
3-3
Water drain
0 to 100 cm above the floor, minimum 5 cm free fall. The water drain
must be located at a lower level than the dialyzer position.
Concentrate supply
0 to -100 mbar; maximum suction height 1 m
with Central Delivery System (option): 0.05 to 2.0 bar
Heat dissipation
Dialysis:
approx. 400 Watt (at an ambient temperature of 20 °C)
Range of operating
temperature
15 °C to 35 °C
Atmospheric pressure
700700 hPa to 1060 hPa
Relative humidity
30 % to 75 %, temporarily 95 %
Stability
5°
IV pole load capacity
Maximum: 5 kg
Maximum load capacity of one hook: 5 kg
O
External connection options
LAN
Interface for the exchange of data.
Electrically isolated by transformer.
Port: RJ 45
RS232
Interface for the exchange of data.
Electrically isolated by optocoupler.
Port: DSUB 9-pin
Caution
Any additional equipment connected to the analog and digital interfaces
of the machine must comply with the applicable EN specifications (e.g.
EN 60950 for data processing equipment and EN 60601 (IEC 601) for
electro-medical equipment).
Apart from this requirement, all configurations must comply with the
system standard EN 60601-1-1 (IEC 601-1-1), or their applicability with
regard to safety has to be proven by a certificate issued by a testing
agency authorized to test the ready-for-use machine.
The connection of additional equipment to the signal input or output
component affects the system configuration and anyone connecting
additional equipment is therefore responsible for compliance with the
system standard EN 60601-1-1 (IEC 601-1-1).
Caution
The external alarm indicators do not relieve the operator of the
obligation to observe the local alarms of the system.
Summary of Contents for 5008
Page 2: ......
Page 6: ...Kapitel 1 Index 1 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 10: ...Chapter 2 Important Information 2 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 22: ...Chapter 3 Specifications 3 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 52: ...Chapter 5 Setup 5 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 60: ...Chapter 5 Setup 5 20 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 66: ...Chapter 6 TSC TMC Maintenance 6 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 74: ...Chapter 6 TSC TMC Maintenance 6 10 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 90: ...Chapter 6 TSC TMC Maintenance 6 26 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 96: ...Chapter 7 Error Messages 7 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 107: ...Chapter 10 Repair Fresenius Medical Care 5008 SM 1 08 04 10 1 10 Repair no contents yet ...
Page 108: ...Chapter 10 Repair 10 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 110: ...Chapter 11 Functional Description 11 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 128: ...Chapter 12 Service Program Option 12 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 130: ...Chapter 12 Service Program Option 12 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 132: ...Chapter 12 Service Program Option 12 6 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 150: ...Chapter 12 Service Program Option 12 24 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 164: ...Chapter 12 Service Program Option 12 38 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 166: ...Chapter 12 Service Program Option 12 40 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 168: ...Chapter 12 Service Program Option 12 42 Fresenius Medical Care 5008 SM 1 08 04 ...