![Fresenius Medical Care 5008 Service Manual Download Page 85](http://html.mh-extra.com/html/fresenius-medical-care/5008/5008_service-manual_2331650085.webp)
Chapter 6: TSC / TMC / Maintenance
Fresenius Medical Care 5008 SM 1/08.04
6-21
TMC
6.4.5
Safety valve tested.
In the service mode, select DIAGNOSTICS. Select BPM from the DIAGNOSTICS
menu.
Basic requirements:
Tube and blood pressure cuff connected.
The blood pressure cuff must be placed on an artificial limb.
Pressure preselection 290 mmHg
Touch the
Calibration test I/O
button.
Once the pressure has reached approx. 290 mmHg, increase the pressure by slowly
pressing the blood pressure cuff. If 320 mmHg ±10 mmHg is exceeded, the cuff must
deflate immediately.
Touch the
Status
button in the Service menu. Touch the
Error memory
button. Touch
the
BPM
button. Check the error memory. Turn the hemodialysis system off and back
on again.
TMC
6.4.6
Blood pressure measurement performed.
Touch the
SYSTEM
button.
Touch the
Blood pressure
button. The blood pressure measurement starts.
After the measurement, check whether the values are plausible.
TMC
6.4.7
Calibration performed.
In the service mode, select DIAGNOSTICS. Select BPM from the DIAGNOSTICS
menu.
Remove tube and blood pressure cuff from the pressure connector.
Connect a rigid metal vessel
(1), a pressure gauge (2), and
an aspirator bulb with drain
valve (3) to the pressure
connector.
Touch the
Calibration test
I/O
button.
Set the appropriate test
pressure using the drain
valve. Wait until the pressure
has stabilized. Check the test
pressure.
250 mmHg/ ±3 mmHg
200 mmHg / ±3 mmHg
150 mmHg / ±3 mmHg
100 mmHg / ±3 mmHg
50 mmHg / ±3 mmHg
6.5
Single
Needle
TSC
6.5.1
Rotor position (Single Needle) checked. Rotor cleaned.
TSC
TMC
MA
No.
Description
1
2
3
Pressure
port
Summary of Contents for 5008
Page 2: ......
Page 6: ...Kapitel 1 Index 1 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 10: ...Chapter 2 Important Information 2 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 22: ...Chapter 3 Specifications 3 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 52: ...Chapter 5 Setup 5 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 60: ...Chapter 5 Setup 5 20 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 66: ...Chapter 6 TSC TMC Maintenance 6 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 74: ...Chapter 6 TSC TMC Maintenance 6 10 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 90: ...Chapter 6 TSC TMC Maintenance 6 26 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 96: ...Chapter 7 Error Messages 7 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 107: ...Chapter 10 Repair Fresenius Medical Care 5008 SM 1 08 04 10 1 10 Repair no contents yet ...
Page 108: ...Chapter 10 Repair 10 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 110: ...Chapter 11 Functional Description 11 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 128: ...Chapter 12 Service Program Option 12 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 130: ...Chapter 12 Service Program Option 12 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 132: ...Chapter 12 Service Program Option 12 6 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 150: ...Chapter 12 Service Program Option 12 24 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 164: ...Chapter 12 Service Program Option 12 38 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 166: ...Chapter 12 Service Program Option 12 40 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 168: ...Chapter 12 Service Program Option 12 42 Fresenius Medical Care 5008 SM 1 08 04 ...