Chapter 4: Installation
4-16
Fresenius Medical Care 5008 SM 1/08.04
Basic requirements:
– Measurement of the protective earth resistance performed.
– Perform the measurement with the hemodialysis system being at operating temperature.
– Dialysate:
Dialysis temperature:
≥
37 °C
Dialysate flow:
≥
300 ml/min
Conductivity:
≥
13 mS/cm
– When performing a direct measurement, the following precautions also must be observed:
The system must be insulated when installed.
All external connections must have been removed from the system.
The line voltage during the measurement will be recorded, as well as the maximum device leakage
current of both mains polarities, scaled to the nominal voltage of the power supply. Maximum device
leakage current: 500
µ
A
Example:
Line voltage during the measurement: 225 V
Device leakage current
for mains polarity 1: 180
µ
A
for mains polarity 2: 120
µ
A
Maximum value of both mains polarities: 180
µ
A
Nominal voltage of power supply: 230 V
Scaled to nominal voltage: 184
µ
A
(180
µ
A: 225 V x 230 V = 184
µ
A
Device leakage current < 500
µ
A: OK
Additional requirements:
If the device leakage current, scaled to the nominal voltage, is higher than 90 % of the admissible alarm
limit (450
µ
A), the last measured value or the first measured value must additionally be considered for
the rating.
If the device leakage current has considerably increased since the last measurement or has continuously
increased since the first measurement (creeping deterioration of the insulation), or if the sum composed
of the current value plus the difference since the last measurement is >500
µ
A, the measurement has
not been completed successfully.
Example 1:
Device leakage current: 470
µ
A
Last measured value: 450
µ
A
470 + (470 – 450) = 470 + 20 = 490; is OK
Example 2:
Device leakage current: 470
µ
A
Last measured value: 390
µ
A
470 + (470 – 390) = 470 + 80 = 550; not passed
10
Check – zero point arterial/venous pressure display and venous clamp
10.1
Zero point arterial pressure display
Basic requirements:
Blood lines inserted, pressure domes coupled, standby operation.
Check the arterial pressure display.
Desired value: –5 mmHg to +5 mmHg
No.
Description
Summary of Contents for 5008
Page 2: ......
Page 6: ...Kapitel 1 Index 1 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 10: ...Chapter 2 Important Information 2 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 22: ...Chapter 3 Specifications 3 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 52: ...Chapter 5 Setup 5 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 60: ...Chapter 5 Setup 5 20 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 66: ...Chapter 6 TSC TMC Maintenance 6 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 74: ...Chapter 6 TSC TMC Maintenance 6 10 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 90: ...Chapter 6 TSC TMC Maintenance 6 26 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 96: ...Chapter 7 Error Messages 7 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 107: ...Chapter 10 Repair Fresenius Medical Care 5008 SM 1 08 04 10 1 10 Repair no contents yet ...
Page 108: ...Chapter 10 Repair 10 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 110: ...Chapter 11 Functional Description 11 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 128: ...Chapter 12 Service Program Option 12 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 130: ...Chapter 12 Service Program Option 12 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 132: ...Chapter 12 Service Program Option 12 6 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 150: ...Chapter 12 Service Program Option 12 24 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 164: ...Chapter 12 Service Program Option 12 38 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 166: ...Chapter 12 Service Program Option 12 40 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 168: ...Chapter 12 Service Program Option 12 42 Fresenius Medical Care 5008 SM 1 08 04 ...