Chapter 3: Specifications
Fresenius Medical Care 5008 SM 1/08.04
3-9
When using ONLINE
plus
™ (option):
Filter life: maximum 100 treatments.
Monitored by the dialysis system and a warning (Filter change) is
displayed. If the warning is ignored, ONLINE
plus
™ will be disabled after
the respective number has been exceeded.
After 90 treatments the number of the remaining treatments will be
displayed in the cleaning programs.
O
OCM (option)
Measuring accuracy of the clearance: ± 6 % standard deviation
Shortest measuring interval: 25 min
Time scale of the display: 10 s
O
ONLINE
plus
™ (option)
Delivery rate: 25 to 600 ml/min (inside line diameter: 8.0 mm)
Resolution: 1 ml/min
Exchange volume: substituate goal 500 l adjustable in relation to
treatment parameters
Accuracy: < 5 % (without lines)
(This specification only applies to the range from 30 to 350 ml/min. With
delivery rates of < 30 ml/min the deviation may be greater.)
Volume counter display: 0.1 to 210 liters
Resolution: 0.1 liter
Spring-loaded rollers, fully occluding, pressure-limited to < 1.3 bar.
(The blood pump design allows manual operation, hand crank in the
rotor, in clockwise direction only.)
Auto sub: The sub rate is determined as a function of:
– UF rate
– Blood flow
– Hematocrit (HCT)
– Total protein (TP)
– Filter performance
O
Single Needle (option)
Blood pump
stop alarm
Single Needle pump
During Single Needle operation 180 seconds.
Stroke volume
10 to 50 ml in increments of 5 ml
External compliance
chamber
50 ml or 60 ml stroke volume
Summary of Contents for 5008
Page 2: ......
Page 6: ...Kapitel 1 Index 1 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 10: ...Chapter 2 Important Information 2 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 22: ...Chapter 3 Specifications 3 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 52: ...Chapter 5 Setup 5 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 60: ...Chapter 5 Setup 5 20 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 66: ...Chapter 6 TSC TMC Maintenance 6 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 74: ...Chapter 6 TSC TMC Maintenance 6 10 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 90: ...Chapter 6 TSC TMC Maintenance 6 26 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 96: ...Chapter 7 Error Messages 7 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 107: ...Chapter 10 Repair Fresenius Medical Care 5008 SM 1 08 04 10 1 10 Repair no contents yet ...
Page 108: ...Chapter 10 Repair 10 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 110: ...Chapter 11 Functional Description 11 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 128: ...Chapter 12 Service Program Option 12 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 130: ...Chapter 12 Service Program Option 12 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 132: ...Chapter 12 Service Program Option 12 6 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 150: ...Chapter 12 Service Program Option 12 24 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 164: ...Chapter 12 Service Program Option 12 38 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 166: ...Chapter 12 Service Program Option 12 40 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 168: ...Chapter 12 Service Program Option 12 42 Fresenius Medical Care 5008 SM 1 08 04 ...