Chapter 6: TSC / TMC / Maintenance
6-28
Fresenius Medical Care 5008 SM 1/08.04
4
Dialysis mode
4.1
Temperature tested with reference instrument.
Desired temperature on temperature display
Difference between system temp. / ref. temp.: –0.5 to
+0.2 °C
System / ref.
____/____
T
4.2
Conductivity tested with reference instrument.
(The bi
b
ag
®
option requires connection of a bi
b
ag
®
.)
Difference between system CD / ref. CD:
≤
±
0.2 mS/cm
System CD / ref. CD
____/____
T
5
Extracorporeal components
5.3
Check of venous clamp performed.
A change in pressure must not exceed the following values within 3 minutes:
Arterial pressure display, maximum change in pressure:
±
5 mmHg
Pressure display of reference measuring instrument, maximum pressure drop: –0.1 bar
T
6
Options
6.3 ONLINEplus
6.3.1
Rotor position (ONLINEplus) checked. Rotor cleaned.
T
6.4 BPM
6.4.4
TMC
Leakage test performed.
Pressure leakage rate: <6 mmHg/min
________
T
6.4.5
TMC
Calibration performed.
Pressure values / tolerance
250 mmHg / ±3 mmHg
200 mmHg / ±3 mmHg
150 mmHg / ±3 mmHg
100 mmHg / ±3 mmHg
50 mmHg / ±3 mmHg
System / ref.
System / ref.
System / ref.
System / ref.
System / ref.
____/____
____/____
____/____
____/____
____/____
T
6.4.6
Safety valve tested.
System emptied at 320 mmHg
±
10 mmHg
________
T
6.4.7
TMC
Blood pressure measurement performed.
Measured values are plausible.
T
6.5 Single
Needle
6.5.1
Rotor position (Single Needle) checked. Rotor cleaned.
T
No.
Description
Measure
ment
value
Summary of Contents for 5008
Page 2: ......
Page 6: ...Kapitel 1 Index 1 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 10: ...Chapter 2 Important Information 2 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 22: ...Chapter 3 Specifications 3 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 52: ...Chapter 5 Setup 5 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 60: ...Chapter 5 Setup 5 20 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 66: ...Chapter 6 TSC TMC Maintenance 6 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 74: ...Chapter 6 TSC TMC Maintenance 6 10 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 90: ...Chapter 6 TSC TMC Maintenance 6 26 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 96: ...Chapter 7 Error Messages 7 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 107: ...Chapter 10 Repair Fresenius Medical Care 5008 SM 1 08 04 10 1 10 Repair no contents yet ...
Page 108: ...Chapter 10 Repair 10 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 110: ...Chapter 11 Functional Description 11 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 128: ...Chapter 12 Service Program Option 12 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 130: ...Chapter 12 Service Program Option 12 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 132: ...Chapter 12 Service Program Option 12 6 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 150: ...Chapter 12 Service Program Option 12 24 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 164: ...Chapter 12 Service Program Option 12 38 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 166: ...Chapter 12 Service Program Option 12 40 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 168: ...Chapter 12 Service Program Option 12 42 Fresenius Medical Care 5008 SM 1 08 04 ...