![Fresenius Medical Care 5008 Service Manual Download Page 65](http://html.mh-extra.com/html/fresenius-medical-care/5008/5008_service-manual_2331650065.webp)
Chapter 6: TSC / TMC / Maintenance
Fresenius Medical Care 5008 SM 1/08.04
6-1
6
TSC / TMC / Maintenance
6.1
Important Information
This chapter includes the Technical Safety Checks (TSC), the
Technical Measurement Checks (TMC) and the Maintenance
Procedures (MA) to be performed.
(Technical Measurement Checks are applicable only to Item 6.4 BPM.)
The Technical Safety Checks (TSC) and the Technical Measurement
Checks (TMC) must be carried out every 2 years (24 months).
Performance of the Technical Safety Checks must be entered in the
Medical Device Register.
The following applies to the technical measurement checks. After
successful completion of the technical measurement checks, the
respective parts of the hemodialysis system must be identified with a
sign (label). This label must, in a unique and traceable manner, specify
the year of the next Technical Measurement Check and the authority or
person having performed the Technical Measurement Check.
Performance of the Maintenance Procedures (MA) is recommended by
the manufacturer. The maintenance procedures must also be carried
out every 2 years (24 months) and ensure smooth operation.
Precautions for working on the system
Assembly, extensions, adjustments, modifications or repairs may only
be carried out by the manufacturer or persons authorized by him.
The activities described in the Technical Manual require the availability
of the necessary technical measuring equipment and accessories.
Respect the following precautions when working on the open system:
Protect the components against ingress of fluids.
Do not touch live parts (e.g. connectors of the power cord or heater).
When repairing and when replacing spare parts, observe the applicable
ESD precautions (e.g. EN 100 015-1).
Summary of Contents for 5008
Page 2: ......
Page 6: ...Kapitel 1 Index 1 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 10: ...Chapter 2 Important Information 2 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 22: ...Chapter 3 Specifications 3 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 52: ...Chapter 5 Setup 5 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 60: ...Chapter 5 Setup 5 20 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 66: ...Chapter 6 TSC TMC Maintenance 6 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 74: ...Chapter 6 TSC TMC Maintenance 6 10 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 90: ...Chapter 6 TSC TMC Maintenance 6 26 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 96: ...Chapter 7 Error Messages 7 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 107: ...Chapter 10 Repair Fresenius Medical Care 5008 SM 1 08 04 10 1 10 Repair no contents yet ...
Page 108: ...Chapter 10 Repair 10 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 110: ...Chapter 11 Functional Description 11 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 128: ...Chapter 12 Service Program Option 12 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 130: ...Chapter 12 Service Program Option 12 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 132: ...Chapter 12 Service Program Option 12 6 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 150: ...Chapter 12 Service Program Option 12 24 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 164: ...Chapter 12 Service Program Option 12 38 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 166: ...Chapter 12 Service Program Option 12 40 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 168: ...Chapter 12 Service Program Option 12 42 Fresenius Medical Care 5008 SM 1 08 04 ...