Chapter 4: Installation
Fresenius Medical Care 5008 SM 1/08.04
4-3
No.
Description
Measure-
ment
value
1
Preparation
1.1
Hemodialysis system without visible shipping damage.
T
1.2
Remove the transport protection for the monitor support arm.
T
Install the IV pole.
T
1.3
Connect the water supply tubing.
T
Connect the drain and the flush tubings.
T
Protect the tubings from slipping out.
T
Standard: Length 3 m, internal diameter 6 mm
Tubing dimensions, adjusted: Length _____ m, internal diameter_____ mm
T
1.4
Connect the CDS tubings and protect them from slipping out.
Apply a shrink tube marking for the CDS tubings.
T
CDS 1
T
CDS 2
T
BIC
T
1.5
Remove the shipping plugs from the overflows.
T
1.6
When turning the hemodialysis system on, perform an audible check of the watchdog alarm.
T
1.7
Rinse out the anti-freeze.
T
1.8
Select the Filter change program.
Connect the filter. DIASAFE
®
plus
/ONLINE
plus
™
In the service mode, delete mandatory disinfection.
Then completely run the rinse program.
T
2
SETUP settings
(Technician's SETUP/Operator SETUP)
2.1
Check the SETUP on the hemodialysis system.
T
3
Check – water inlet flow / adjustment – degassing
3.1
Check MaxWaterFlow.
Desired value: 1300 ml/min to 1550 ml/min
________
Corr.:
T
Yes
T
No
3.2
Perform the degassing adjustment.
T
4
Check – dialysate flow
4.1
Check flow at 800 ml/min.
Desired value: 770 ml/min to 830 ml/min
________
T
Summary of Contents for 5008
Page 2: ......
Page 6: ...Kapitel 1 Index 1 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 10: ...Chapter 2 Important Information 2 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 22: ...Chapter 3 Specifications 3 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 52: ...Chapter 5 Setup 5 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 60: ...Chapter 5 Setup 5 20 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 66: ...Chapter 6 TSC TMC Maintenance 6 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 74: ...Chapter 6 TSC TMC Maintenance 6 10 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 90: ...Chapter 6 TSC TMC Maintenance 6 26 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 96: ...Chapter 7 Error Messages 7 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 107: ...Chapter 10 Repair Fresenius Medical Care 5008 SM 1 08 04 10 1 10 Repair no contents yet ...
Page 108: ...Chapter 10 Repair 10 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 110: ...Chapter 11 Functional Description 11 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 128: ...Chapter 12 Service Program Option 12 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 130: ...Chapter 12 Service Program Option 12 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 132: ...Chapter 12 Service Program Option 12 6 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 150: ...Chapter 12 Service Program Option 12 24 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 164: ...Chapter 12 Service Program Option 12 38 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 166: ...Chapter 12 Service Program Option 12 40 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 168: ...Chapter 12 Service Program Option 12 42 Fresenius Medical Care 5008 SM 1 08 04 ...