MA
4.3
Dialysate pressure checked.
In the service mode, select CALIBRATE.
Basic requirements:
The hemodialysis system
must be closed.
The reference measuring instrument must be placed at the bottommost position of the
IV pole.
Dialyzer couplings must be connected to the reference instrument.
Flow on until dialysate lines and reference measuring instrument are free from air. Then
flow off.
4.3.1
Zero point S03/S07
Reference measuring instrument: 0 bar
Open the vent valve (UMED).
Using a syringe (filled with fluid) set a value of 0 bar, via the vent valve.
Check S03.
Desired value: +16 mbar to +76 mbar
Check S07.
Desired value: +16 mbar to +76 mbar
4.3.2
Slope S03/S07 (+)
Reference measuring instrument: +533 mbar (
±
26 mbar)
Using a syringe (filled with fluid) set a value of +533 bar, via the vent valve.
Check S03.
Desired value: S03 = Display of reference measuring inst (+16 mbar to
+76 mbar)
Check S07.
Desired value: S07 = Display of reference measuring inst (+16 mbar to
+76 mbar)
4.3.3
Slope S03/S07 (–)
Reference measuring instrument: –533 mbar (
±
26 mbar)
Using a syringe (filled with fluid) set a value of –533 bar, via the vent valve.
Check S03.
Desired value: S03 = Display of reference measuring inst (+16 mbar to
+76 mbar)
Check S07.
Desired value: S07 = Display of reference measuring inst (+16 mbar to
+76 mbar)
5
Extracorporeal components
MA
5.1
Zero point of arterial pressure display checked (standby operation)
Desired value: –5 mmHg to +5 mmHg
MA
5.2
Zero point of venous pressure display checked (standby operation)
Desired value: –5 mmHg to +5 mmHg
TSC
TMC
MA
No.
Description
Summary of Contents for 5008
Page 2: ......
Page 6: ...Kapitel 1 Index 1 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 10: ...Chapter 2 Important Information 2 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 22: ...Chapter 3 Specifications 3 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 52: ...Chapter 5 Setup 5 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 60: ...Chapter 5 Setup 5 20 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 66: ...Chapter 6 TSC TMC Maintenance 6 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 74: ...Chapter 6 TSC TMC Maintenance 6 10 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 90: ...Chapter 6 TSC TMC Maintenance 6 26 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 96: ...Chapter 7 Error Messages 7 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 107: ...Chapter 10 Repair Fresenius Medical Care 5008 SM 1 08 04 10 1 10 Repair no contents yet ...
Page 108: ...Chapter 10 Repair 10 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 110: ...Chapter 11 Functional Description 11 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 128: ...Chapter 12 Service Program Option 12 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 130: ...Chapter 12 Service Program Option 12 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 132: ...Chapter 12 Service Program Option 12 6 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 150: ...Chapter 12 Service Program Option 12 24 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 164: ...Chapter 12 Service Program Option 12 38 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 166: ...Chapter 12 Service Program Option 12 40 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 168: ...Chapter 12 Service Program Option 12 42 Fresenius Medical Care 5008 SM 1 08 04 ...