Chapter 9: Calibration / Adjustment
9-4
Fresenius Medical Care 5008 SM 1/08.04
O
Flow pump (A03)
Select the
Flow pump (A03)
field. Touch the
Start calibration
button.
The following message will be displayed:
Adjustment of relief valve A03.
between 2550 and 2650 mbar Pressure S15: XXXX mbar.
If the
pressure is within the range, touch the
Confirm
button
.
The CALIBRATE SCREEN will be displayed. The flow pump (A03) is
now calibrated.
If the pressure limitation is not within the desired range, it must be
adjusted using A03.
O
Temperature (PT7/ PT8/ PT9)
Select the
Temperature (PT7/ PT8/ PT9)
field. Touch the
Start
calibration
button.
Measure the reference temperature with an external measuring
instrument.
Basic requirements:
The hemodialysis system
must be closed.
Flow on, response time approx. 10 minutes
Select the
Ref. temp.
field. Enter the reference temperature. Touch the
OK
button to confirm the value entered. Touch the
Accept value
button.
The CALIBRATE SCREEN will be displayed.
A03
Summary of Contents for 5008
Page 2: ......
Page 6: ...Kapitel 1 Index 1 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 10: ...Chapter 2 Important Information 2 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 22: ...Chapter 3 Specifications 3 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 52: ...Chapter 5 Setup 5 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 60: ...Chapter 5 Setup 5 20 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 66: ...Chapter 6 TSC TMC Maintenance 6 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 74: ...Chapter 6 TSC TMC Maintenance 6 10 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 90: ...Chapter 6 TSC TMC Maintenance 6 26 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 96: ...Chapter 7 Error Messages 7 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 107: ...Chapter 10 Repair Fresenius Medical Care 5008 SM 1 08 04 10 1 10 Repair no contents yet ...
Page 108: ...Chapter 10 Repair 10 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 110: ...Chapter 11 Functional Description 11 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 128: ...Chapter 12 Service Program Option 12 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 130: ...Chapter 12 Service Program Option 12 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 132: ...Chapter 12 Service Program Option 12 6 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 150: ...Chapter 12 Service Program Option 12 24 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 164: ...Chapter 12 Service Program Option 12 38 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 166: ...Chapter 12 Service Program Option 12 40 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 168: ...Chapter 12 Service Program Option 12 42 Fresenius Medical Care 5008 SM 1 08 04 ...