Chapter 4: Installation
4-10
Fresenius Medical Care 5008 SM 1/08.04
1.5
Remove the shipping plugs from the overflows.
1. Vent (water inlet chamber)
2. Vent (mixing chamber)
1.6
When turning the hemodialysis system on, perform an audible check of the watchdog alarm.
1.7
Rinse out the anti-freeze.
1.8
Select the Filter change program.
Connect the filter. DIASAFE
®
plus
/ONLINE
plus
™
In the service mode, delete mandatory disinfection.
Then completely run the rinse program.
2
SETUP settings
(Technician's SETUP/Operator SETUP)
2.1
Check the SETUP on the hemodialysis system.
Make the appropriate settings for the respective hospital, if necessary.
3
Check – water inlet flow / adjustment – degassing
3.1
In the service mode, select FLOW DIAGRAM.
Basic requirements:
The hemodialysis system must be closed.
Flow on.
Check MaxWaterFlow.
Desired value: 1300 ml/min to 1550 ml/min
Use A04 for making corrections, if necessary.
(If it is not possible to set a water inlet flow
≥
1300ml/min, it will not always be possible to achieve the
dialysate flow of 1000ml/min.)
3.2
In the service mode, select CALIBRATE.
Basic requirements:
Flow on.
Perform the degassing adjustment.
Touch the
Degassing (A01/P01)
button.
No.
Description
1
2
Summary of Contents for 5008
Page 2: ......
Page 6: ...Kapitel 1 Index 1 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 10: ...Chapter 2 Important Information 2 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 22: ...Chapter 3 Specifications 3 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 52: ...Chapter 5 Setup 5 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 60: ...Chapter 5 Setup 5 20 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 66: ...Chapter 6 TSC TMC Maintenance 6 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 74: ...Chapter 6 TSC TMC Maintenance 6 10 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 90: ...Chapter 6 TSC TMC Maintenance 6 26 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 96: ...Chapter 7 Error Messages 7 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 107: ...Chapter 10 Repair Fresenius Medical Care 5008 SM 1 08 04 10 1 10 Repair no contents yet ...
Page 108: ...Chapter 10 Repair 10 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 110: ...Chapter 11 Functional Description 11 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 128: ...Chapter 12 Service Program Option 12 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 130: ...Chapter 12 Service Program Option 12 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 132: ...Chapter 12 Service Program Option 12 6 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 150: ...Chapter 12 Service Program Option 12 24 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 164: ...Chapter 12 Service Program Option 12 38 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 166: ...Chapter 12 Service Program Option 12 40 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 168: ...Chapter 12 Service Program Option 12 42 Fresenius Medical Care 5008 SM 1 08 04 ...