Chapter 4: Installation
Fresenius Medical Care 5008 SM 1/08.04
4-1
4
Installation
4.1
Preface
Instructions for all technicians who are authorized to commission
our hemodialysis systems.
We, as manufacturers, permanently aim at delivering systems of
highest quality.
To reach this aim, we need your support.
Please commission our hemodialysis systems by uniformly using the
enclosed "initial start-up report" and enter the values determined in the
columns provided.
The following is applicable:
Corrections are necessary only if the measured values are outside
of the tolerances specified!
We will then evaluate the initial start-up reports, which will enable us to
monitor the quality of our systems on their delivery.
After initial start-up, please asap send – by mail or by fax – the
completed form (Initial Start-Up Report) back to the following address:
Fresenius Medical Care
Deutschland GmbH
Werk Schweinfurt
Herrn Alfred Laus, Abt. BM
Hafenstraße 9
97424 Schweinfurt
Fax: 09721/ 678450
Thank you very much for your help!
4.2
Important Information on Initial Start-Up
This technical document is intended for initial start-up only. It is not
intended for restarting hemodialysis systems that have been shut down
or have been put out of service temporarily.
The initial start-up must be performed by the Technical Service of
Fresenius Medical Care or a person authorized by them!
Any information on initial start-up and the specifications in the Operating
Instructions must be observed.
When bringing the hemodialysis system from a cooler to a warmer
room, allow approx. 2 hours for the system to adjust to the ambient
temperature before turning the unit on.
Summary of Contents for 5008
Page 2: ......
Page 6: ...Kapitel 1 Index 1 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 10: ...Chapter 2 Important Information 2 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 22: ...Chapter 3 Specifications 3 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 52: ...Chapter 5 Setup 5 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 60: ...Chapter 5 Setup 5 20 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 66: ...Chapter 6 TSC TMC Maintenance 6 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 74: ...Chapter 6 TSC TMC Maintenance 6 10 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 90: ...Chapter 6 TSC TMC Maintenance 6 26 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 96: ...Chapter 7 Error Messages 7 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 107: ...Chapter 10 Repair Fresenius Medical Care 5008 SM 1 08 04 10 1 10 Repair no contents yet ...
Page 108: ...Chapter 10 Repair 10 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 110: ...Chapter 11 Functional Description 11 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 128: ...Chapter 12 Service Program Option 12 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 130: ...Chapter 12 Service Program Option 12 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 132: ...Chapter 12 Service Program Option 12 6 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 150: ...Chapter 12 Service Program Option 12 24 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 164: ...Chapter 12 Service Program Option 12 38 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 166: ...Chapter 12 Service Program Option 12 40 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 168: ...Chapter 12 Service Program Option 12 42 Fresenius Medical Care 5008 SM 1 08 04 ...