Chapter 5: Setup
Fresenius Medical Care 5008 SM 1/08.04
5-13
5.2
Technician's SETUP
Selecting the technician's SETUP
System turned power on.
Insert the technician's card.
Touch the
SYSTEM
button.
Touch the
SERVICE
button in the SYSTEM screen.
Touch the
SETUP
button on the SERVICE SCREEN.
How to use the OPERATOR SETUP
Save with OK.
Touch the logo to select default values.
O
Hydraulics settings
O
EBM settings
Submenu
Default value
Value range
Resolution
Selectable
options
Machine options
Machine options
DIASAFE
®
plus
or
ONLINE
plus
™
(not adjustable)
–
–
–
Filter 1
–
–
–
Filter not present
Present
Filter 2
–
–
–
Filter not present
Present
Water inlet tube
Length
3.0 m
1.0–5.0 m
0.1 m
–
Internal diameter
6 mm
3–20 mm
1 mm
–
Submenu
Default value
Value range
Resolution
Selectable
options
Machine options
Motor type
Premotec
–
–
Premotec
Papst
others
Summary of Contents for 5008
Page 2: ......
Page 6: ...Kapitel 1 Index 1 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 10: ...Chapter 2 Important Information 2 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 22: ...Chapter 3 Specifications 3 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 52: ...Chapter 5 Setup 5 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 60: ...Chapter 5 Setup 5 20 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 66: ...Chapter 6 TSC TMC Maintenance 6 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 74: ...Chapter 6 TSC TMC Maintenance 6 10 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 90: ...Chapter 6 TSC TMC Maintenance 6 26 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 96: ...Chapter 7 Error Messages 7 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 107: ...Chapter 10 Repair Fresenius Medical Care 5008 SM 1 08 04 10 1 10 Repair no contents yet ...
Page 108: ...Chapter 10 Repair 10 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 110: ...Chapter 11 Functional Description 11 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 128: ...Chapter 12 Service Program Option 12 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 130: ...Chapter 12 Service Program Option 12 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 132: ...Chapter 12 Service Program Option 12 6 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 150: ...Chapter 12 Service Program Option 12 24 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 164: ...Chapter 12 Service Program Option 12 38 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 166: ...Chapter 12 Service Program Option 12 40 Fresenius Medical Care 5008 SM 1 08 04 ...
Page 168: ...Chapter 12 Service Program Option 12 42 Fresenius Medical Care 5008 SM 1 08 04 ...