xii TIMS MVP
™
Setup and Configuration Guide
Table 6
–
Recommended separation distances between portable
and mobile RF communications equipment and the ME
EQUIPMENT or ME SYSTEM
–
for ME EQUIPMENT and ME SYSTEMS
that are not LIFE-SUPPORTING
(see 5.2.2.2)
Recommended separation distances between portable and mobile RF communications
equipment and the TIMS MVP
™
System
The TIMS MVP
™
System is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the TIMS MVP
™
System
can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the TIMS MVP
™
System
as recommended below, according to the maximum output power of the communications
equipment.
WARNING:
Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any
part of the TIMS MVP
™
System, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
Rated maximum
output power of
transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
0,01
0.12
0.12
0.23
0,1
0,37
0,37
0,74
1
1,17
1,17
2,33
10
3,70
3,70
7,37
100
11,70
11,70
23,30
For transmitters rated at a maximum output power not listed above, the recommended
separation distance
d
in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Summary of Contents for TIMS 2000 EN
Page 1: ...060000 91 Rev 1...