51
Trendline 3
calibration and want to use the balancing function on a different Detector in the
meantime.
To activate the balancing function on the Detector, start the Trendline software.
Connect the Detector to the PC using the serial cable. Plug the dongle into a free
USB port. Switch on the Detector and
select
the Trendline menu item
Detector
>
balancing activation
. The following window is now displayed:
The serial number of the connected Detector is displayed at the top. The dongle
that has been identified is displayed at the bottom. The identified version should
start with "3" in the case of Detector III. The serial number of the Detector for
which this dongle has already been used or, if the dongle has not yet been used,
Empty
is displayed in the
Used for Detector:
column. Now select the required
dongle from the
Type
column (should be displayed as "Empty" in the "Used for
Detector" column). You can now enable the balancing functionality via
Activate
balancing for connected Detector
. A new line
Balancing
should now be visible
and the serial number of the Detector is displayed in the dongle line.
Please note the following: