
222
Activate constant supply of the sensor:
Select
Sensor supply
>
Always on
and confirm the warning for passive
sensors.
The sensor is initialized. If an error occurs during the sensor initialization, the
Detector will signal a sensor error and retain the
If required
setting.
Deactivate constant supply of the sensor:
Select
Sensor supply
>
If required
.
Power down time
Here you can configure when the device should switch off automatically after the
last action.
Select
Power down time
.
Press the Enter button
and
set the power down time in minutes.
"0" deactivates this function. In order to save battery running time, set the power
down time as short as possible.
Bias voltages
For ICP sensors you can define the minimum and maximum value of the bias
voltage. Minimum bias voltage must be at least 3, maximum bias voltage must
not be greater than 17. The difference between the two values must not be less
than 10. The default setting is min. 5 V and max. 17 V.
This values will be used for the free measurements "ISO 10816" and "Headset"
as thresholds for the sensor test.
RFID settings
This menu item is only visible if the Detector has an RFID reader.
Select whether the Detector should confirm successful import of an RFID tag. The
following settings are available: optical, acoustic, both.
6.6 Measuring procedure
During a measuring round the sensor signals are recorded at all measuring
points and the characteristic values calculated. The measuring points can be
measured in random order.
Before you perform a round of measurements with your Detector, you should
You should mark the measuring position, where the sensor should be
mounted for the measurement, in a suitable way. (only then will you get
comparable results usable for trend analysis), and