Instruction Manual MM850432 ed. 05c
90
Direct test of Pressure Transmitters at actual programmed test points or with an automatic calculation of actual errors with inlet pressure in an acceptable deviation band from the
Test Point level.
NOTE
P
ARAMETERS OF AN OFF
-
LINE PROCEDURE MUST BE FIRSTLY TRANSFERRED ON THE
M
ICRO
C
AL
:
SEE PAR
.6.2
•
Press
[
CAL PROC
] and select the proper Tag from Tag item (in this example, the file is associated to the procedure is “Tag1”)
SELECT Procedure
TagSel []
Results [Found]
Display procedure
Display results
Run procedure
Press [ ] to SET
File selection
Tag 1
Tag 2
Tag 3
Press [ ] to SET
SELECT Procedure
TagSel [Tag1]
Results [Found]
Display procedure
Display results
Run procedure
Press [ ] to SET
•
Once loaded data calibration, it is possible to check them selecting “
Display procedure
”. One or more pages appear, interchangeable by pressing [
ENTER
]:
SELECT Procedure
Tag [Tag1]
Results [Found]
Display procedure
Display results
Run procedure
Press [ ] to SET
Tag1
My Tag 1
Description 1
Description 2
Ref. Limit
±
(0.01%+2.0mV)
Press [ENTER]
Tag1
Actual limit
±
(0.025%+2.0mV)
Press [ENTER]
Tag1
Test Points:
0.0000 | 3.0000
1.0000 | 4.0000
2.0000 | 5.0000
2.5000 |
Press [ENTER]
Even from “Select procedure”, it is possible to set the procedure as “
As found
” or “
As Left
”, it means to state whether the instrument is as found or as left after a calibration