13 MAINTENANCE
The calibrator has been developed and manufactured using high quality components. A correct, systematic maintenance will prevent damage and will increase substantially the life of
the instrument, giving excellent results. Anomalous conditions that require extraordinary maintenance interventions are promptly displayed, either clearly or with an information code.
13.1 Error
messages
MicroCal 20
performs the auto-diagnostic procedure when switched on. When measuring it checks also for anomalous conditions. If the calibrator founds trouble, it displays a
message on LCD. Sometimes, these messages correspond to an unrecoverable problem and it is necessary to return the instrument to Eurotron for maintenance. Frequently these
messages are referred to a recoverable trouble and can be solved by you.
The following table shows the possible
MicroCal 20
error messages:
Error message
Problem
Solutions
Over
Flashes instead of the related engineering unit and it
indicates an overcome of the full scale
Insert a value below to the full scale
Under
Flashes instead of the related engineering unit and it
indicates an overcome of the zero.
Insert a value above the zero
Open
Indicates that a Tc is badly connected or interrupted
Check the Tc connections
Unsafe
Indicates that a protection occurred
The channel is locked until a new setting of the input
Battery Low
Low batteries (on blank page)
Charge the batteries
File not found
in <
directory
>
A file is missing for calibration procedure or
linearization
Look on your PC for that file
File exists or system
full
This error can occur during the creation of a Log file
Change filename or free some space into the memory
Log active don
’
t modify
The Log is active, it is impossible to modify parameters
Authorized persons
only
Attempt to enter user level “factory” not allowed
Choose another user level
Danger!!
Overcurrent on Ch1
(+ acoustic signal)
Channel 1 is under a too strong current that can
damage permanently the instrument
The instrument enters the safe mode. Use less current or disconnect the source
Danger!!
Overpressure on Pl
(+ acoustic signal)
Channel Pl is under a too strong pressure that can
damage permanently the sensor
Use less pressure or disconnect the source
Replay active not
accepted
A forbidden operation was done during the Replay
function
Wait for the end of simulation (Replay) or stop it