15.3.2 Battery
maintenance
Do not store the calibrator with discharged or partial discharged batteries. Always recharge the batteries after you have finished your job. The batteries are fully charged at 7.2 Volt
(NiMH) batteries are nearly discharged at 6.8 Volt.
During operation a fully battery symbol “
≠
” will be displayed on the display. This symbol means that the batteries are completely full. When the batteries will be discharged the symbol
“
–
” will appear and the instrument still has about 20 minutes operation capability to end the running analysis.
The battery symbol indicates that a full charge is required. Use only the dedicated battery charger supplied by Eurotron together the instrument.
CAUTION:
O
LD BATTERIES CAN LEAK AND CAUSE CORROSION
.
N
EVER LEAVE RUN DOWN BATTERIES IN THE INSTRUMENT
15.3.3
Battery pack replacement
The batteries are a part of the instrument housing. Open the calibrator (ONLY IN SAFE AREAS) by releasing all visible screws at the back and inside the battery compartment. Lift
gently the housing lower half off and disconnect the cable with the white plug. Proceed reverse to install the new pack. Battery pack replacement must be carried out by qualified
technicians only. Ask your distributor. No recalibration is required.