Instruction Manual MM850432 ed. 05c
84
If you want to insert polynomial coefficients for a RTD, you have to select “
Set RTD parameters
” and insert proper numerical values.
Example
: for a Pt100 in the range from 0°C to 850°C, the parameters are the following: R0=100 (resistance in ohm at 0°C), A=3.90802
·10
-3
°C
-1
, B=-5.802 ·10
-7
°C
-2
, C=0
After defining linearization points, it is possible to show the related graphic through the icon
or to save the related file (“mcl” extension) with
. Besides, you can load a
linearization file for eventual changes with
.
To transfer a linearization file to the instrument, it must appear in the Lin folder. In order to do that it is possible just to drag and drop it or to click on the 3
rd
icon of Explorer “
Send file
to instrument
” and select the proper file (“mcl”).
•
Run
LinMan
from
Tools
Menu: