Instruction Manual MM850432 ed. 05
65
OnOff
: run/stop the test
Channel
: set the pressure input for the test
Wait time:
seconds before starting the leak test procedure
Leak Time:
leak test interval (seconds)
•
Select
"
OnOff
" to start the test
The set waiting time will be start and the following pages will appear
Leak Test
Waiting Time 10
Actual = 100.00
Press ESC to abort
the test start and the set leak time will countdown;
Leak Test
Leak Time 60
Initial = 100.00
Actual = 99.50
Press ESC to abort
at the end of the test the following page will appear
Leak Test
Initial = 100.00
Final = 99.50
Leak = 0.50
Leak time = 60
Leak rate = 0.50
Press enter to store
•
Press
[
ENTER
] to store the results of the test in a log file.
5.14.1
Leak Test: file
Results of leak test can be stored in a text file with progressive numeration: “Leak test N.txt”, where N is the number of files.
This file is stored in the instrument memory in the “log” folder accessible with Utilities Manager. See chap. 6.
This txt file has a structure like that in the picture: