Instruction Manual MM850432 ed. 05c
52
5.5.13 Frequency/Pulse:
measure
IMPORTANT
M
EASURE OF FREQUENCY OR PULSES IS POSSIBLE ONLY ON CHANNEL
1
•
Press the [
SELECT
] key until “
SELECT CH1
” is shown
SELECT Ch1
Thermocouple
RTD thermometer
Current
Voltage
Resistance
Press < > to SET
Select “
Frequency/Pulse
”
Frequency/Pulse
Frequency
Pulse
Press < > to SET
select “
Frequency
” o “
Pulse
”
•
Press the [ ] key to select the option. Press the [
ESC
] key to return in measure mode without changes
If you select Frequency as a input:
If you select Pulse as a input:
Frequency
Range
Trigger level [2]
Press < > to SET
•
Select
“
Trigger level
” to define the threshold level (Volt)
•
Select “
Range
” to define the frequency level
Frequency
200Hz
2kHz
20kHz
Press < > to SET
•
Press the [ ] key to select the option. Press the [
ESC
] key to return in
measure mode without changes
Se scegliete “
Pulse
” come input, lo strumento mostra:
Pulse
Measure
Trigger level [2]
Time
Mode [One shot]
Press < > to SET
•
Select
“
Trigger level
” to define the threshold level (Volt)
•
Define the time, in seconds, used to perform the measure, with “
Time
”
•
Select the mode for the measure of pulses: “
One shot
” or “
Continuous
”, depending
on which kind of measure you want: over a single time range or continuously.
•
Select Measure to start the measure. The instrument comes back automatically to
the measuring page and channel 1 measures a value expressed in p/t