Instruction Manual MM850432 ed. 05
63
•
If the sequencer is programmed in “AUTO” mode, press the [
HOLD
] key to pause and to continue the procedure.
•
If the sequencer is programmed in “MANUAL” mode, press the [
HOLD
] key to go to next step.
5.13 Switch
Test
Verifying functions are available for measurement’s instruments of temperature or pressure.
•
Press
[
MENU
], the following page is shown:
MENU
Configuration
Unit
Ramp-Memory scan
Data logger
Advanced
Press < > to SET
Select “
Advanced
”
Advanced
Switch Test
LeakTest
HART
Trx simulator
Graph
Press < > to SET
Select “
Switch Test
”
Switch Test
Pressure
Temperature
Switch [Ext]
Press < > to SET
You can also set the contact input for the switch test by selecting it between the "
external
", "
ch1
" or "
ch2
" options:
Select "
Switch…
"
Switch Test
Pressure
Temperature
Switch [Ext]
Press < > to SET
and the select the contact input to set it
Switch
Ext
Ch1
Ch2
Press < > to SET
NOTE:
THE EXTERNAL CONTACT INPUT
[EXT]
IS NOT AVAILABLE ON
IS
MODELS