Instruction Manual MM850432 ed. 05c
70
•
Connect the instrument to PC through the serial communication cable (BB530203) and turn it on and enter the Firmware Upgrade mode by pressing [
MENU
]+[
ON
]
•
Select the right COM port (
Serial port
)
•
Select the same communication velocity (expressed in Baud) either on PC (
Baud rate
) and on the instrument
ATTENTION
S
ET THE SAME BAUD RATE EITHER ON
PC
AND THE INSTRUMENT
,
OTHERWISE NO COMMUNICATION HAPPENS
.
I
N CASE OF COMMUNICATION PROBLEMS
,
TRY TO DECREASE THE BAUD RATE
.
•
Insert in the field “Filename” the position where the file with “HFW” extension is, it means the firmware that has to be transferred to the instrument.
•
Press “Program” and wait the complete data transfer
•
Press “END” to quit the program
NOTE:
EVENTUALLY
,
PRESS
“R
EAD ME
”
TO READ AN HELP ONLINE
.
ATTENTION
I
N CASE OF FAILED PROCEDURE
,
TURN THE INSTRUMENT OFF BY PRESSING
[ON/OFF]
FOR ABOUT
10
SECS AND REPEAT THE OPERATION
.