Instruction Manual MM850432 ed. 05
77
8.1.2
Data logger: data simulation (Replay)
Replay function allows recalling stored data in a log file.
•
Press
[
MENU
]:
MENU
Configuration
Unit
Ramp-Memory scan
Data logger
Advanced
Press < > to SET
Select “
Data logger
”
Data logger
On/Off
Source
Time
Name [MyLog]
Replay
Press < > to SET
select a desired log file specifying the
Name
•
Select
Replay
:
Replay
On/Off [off]
Column [1]
Press < > to SET
select the proper number of column, it means the signal to simulate among those previously stored. Indeed, a log file is divided in many data columns depending on how
many fields (Sources) are active during storing phase.
•
Select the “
On/Off
” key to run the procedure
NOTE:
SET
C
HANNEL
2
AS
OUT
IF YOU WANT TO GENERATE A SIGNAL
The display shows the following page:
15.2000
Ch1 mA
Ch2 Out TcK Rji °C
100.02
R
The symbol “
R