Instruction Manual MM850432 ed. 05
101
Select “
Function
” parameter and change it to “linear”.
Select “
Decimals
” parameter and change it to “2” decimal digits.
Set the CH1 in input mode and configure it as XScale (e.g. X1).
SELECT Ch1
Thermocouple
RTD thermometer
Current
Voltage
Resistance
Press < > to SET
press the [ ] key until the “
Process Function
” option is highlighted:
SELECT Ch1
Voltage
Resistance
Frequency/Pulse
Process Function
Rounding: 0.0000
Press < > to SET
Process Function
XScale
LinMan [ ]
Press < > to SET
Xscale
X1
X2
X3
X4
Set X1
Press < > to SET
100.13
Ch1 X1 °C
Ch2 Out TcK Rji °C
100
.
00
Change the output value by using the 6 [
] keys (You can also use the calculator mode to insert the value).
Read on the upper display the transmitter output.