Instruction Manual MM850432 ed. 05c
40
•
Press the [ ] and [ ] keys to highlight the “
Thermocouple
” parameter.
•
Press the [ ] key to select the option. Press the [
ESC
] key to return in measure mode without changes.
Thermocouple
Rji
Rje
Rjr
Press < > to SET
•
Press the [ ] and [ ] keys to highlight the reference joint (Rj) mode.
Rji
Internal reference. The temperature is automatically compensated by the internal sensor.
Rje
External reference. The temperature is compensated with the value programmed in “Menu – Unit – Temperature – Rj Ch1 ext” menu.
Rjr
Remote reference. The temperature is compensated using the channel 2 as RTD input. This option is available only on
MicroCal 20DPC XP
and
MicroCal 20T
models.
•
Press the [ ] key. Press the [
ESC
] key to return in measure mode without changes.
Rji
TcK
TcJ
TcT
TcR
TcS
Press < > to SET
•
Press the [ ] and [ ] keys to highlight the requested type of thermocouple.
•
Press the [ ] key to confirm the selection and going in measuring mode. Press the [
ESC
] key to return in measure mode without changes.
5.5.2
External Cold Joint Reference Setting
•
Press the [