Instruction Manual MM850432 ed. 05
87
However, MicroCal 20 allows storing inside itself all parameters needed for calibration procedures: in this case, we speak about off-line procedures as the MicroCal is disconnected
from PC and it can be moved on the field to perform the required procedures. About procedures off-line, see the following chapter. Anyway, the CalpMan 2007 manual is still useful for
the definition of the procedure.
8.2.1 Transfer
calibration
parameters
Use CalpMan 2007 to define the calibration procedure: there is no differences between the definition of a calibration procedure on-line and off-line. Follow the instructions shipped with
the software.
ATTENTION
A
S A PROCEDURE IS TRANSFERRED ON THE CALIBRATOR
,
PAY ATTENTION TO THE DEFINITION OF
COM
PORTS FOR ACTUAL AND REFERENCE VALUES
.
•
Select a proper Tag from CalpMan 2007, like the example:
•
To transfer data on the calibrator, you need an
Upload
of information. Hence, be sure that the instrument is switched on and properly connected to PC through serial port.