Instruction Manual MM850432 ed. 05
41
MENU
Configuration
Unit
Ramp-Memory scan
Data logger
Advanced
Press < > to SET
•
Press the [ ] and [ ] keys to highlight the “
Unit
” option.
•
Press the [ ] key to confirm the selection. Press the [
ESC
] key to return in measure mode without changes.
Unit
Temperature
Pressure [mbar]
Xscale
Press < > to SET
Select the “
Temperature
” option.
Temperature
Unit [°C]
Rj Ch1 ext [0.0]
Rj Ch2 ext [0.0]
Scale [ITS 90]
Press < > to SET
•
Press the [ ] and [ ] keys to highlight the “
Rj Ch1 ext
” option for channel 1 setting or highlight the “
Rj Ch2 ext
” option for channel 2 setting.
•
Press the [ ] key to confirm the selection.
Rj Ch1 ext
20.0
Use numeric keys
[ENTER] to confirm
Use the numeric pad (red keys) to insert the temperature value (e.g. 25.0°C)
Rj Ch1 ext
0.0
25.0
•
Press
[
ENTER
] to confirm or [
] to go to the upper menu level without changes.
•
Press the [
ESC
] key to return in measure mode without changes.
5.5.3 Temperature
with
RTDs
The calibrator can support 14 standard of resistance thermometers. See table on Technical Specifications chapter.
Two, Three, or Four wire RTD connections can be performed for high accuracy measurements. Refer to the electrical connection chapter.