Instruction Manual MM850432 ed. 05c
60
5.9 Data
Logging
See par.6.1.1 as this function is strictly related to communication with PC
5.10 Graph
The graphic display allows showing the real time trend for two channels.
The procedure uses channels configuration. Before run the procedure the channels to be graph has to be set.
•
Press the [
MENU
] key. The following page is shown:
MENU
Configuration
Unit
Ramp-Memory scan
Data logger
Advanced
Press < > to SET
Select the “
Advanced
” option
Advanced
Switch Test
Leak Test
HART
Trx Simulator
Graph
Press < > to SET
Select the “
Graph
” option
Graph
Source [Ch1]
In Low [0.00]
In High [100.00]
Time
Mode [Line]
Press < > to SET
Source
: Select the graph channel source. Available settings are: CHP (if internal and/or external pressure sensor are installed), CH1, CH2.
In Low
: Set the graph Low scale
In High
: Set the graph full scale
Time
: Set the sampling time (expressed in min. and sec.)
Mode
: set the mode between the options "
Dot
" and "
Line
"
•
To display the graphic, select the graphic display mode by pressing the [
DISPLAY
] key and selecting the “
Graph
” item.