Instruction Manual MM850432 ed. 05
17
IN Frequency/Pulse
RANGE
RESOL.
ACCURACY
Plus model
1 to 200 Hz
0.001 Hz ±(0.005% rdg + 0.001 Hz)
1 to 2000 Hz
0.01 Hz
±(0.005% rdg + 0.01 Hz)
1 to 20000 Hz
0.1 Hz
±(0.005% rdg + 0.1 Hz)
Input impedance:
Pressure (option)
Pressure media: AISI 316 SS compatible fluids (water, gas, and oil)
Temperature compensation: Automatic with built-in calibration matrix.
Engineering units: mbar, bar, Pa, hPa, kPa, MPa, kg/cm2, kg/m2, psi, mmH2O,
cmH2O, mH2O, Torr, atm, lb/ft2, inH2O, ftH2O, inH2O@4°C, ftH2O@4°C, mmHg,
cmHg, mHg, inHg, programmable.
Internal sensors
RANGE
RESOL.
ACCURACY
-100 to 100 mbar
0.001mbar
±0.25 % F.S.
-500 to 500 mbar
0. 01mbar
±0.25 % F.S.
-0.95 to 2 bar
0.01mbar
±0.25 % F.S.
0 to 2 bar (ABS)
0.01mbar
±0.25 % F.S.
-0. 95 to 7 bar
0.1mbar
±0.25 % F.S.
-0. 95 to 20 bar
0.1mbar
±0.25 % F.S.
0 to 20 bar (ABS)
0.1mbar
±0.25 % F.S.
Overpressure: 125% F.S.
Port: 1/8 BSPF (female)
External sensors
PART No.
RANGE
RESOL.
ACCURACY
BB480009
-100 to 100 mbar
0.001 mbar ±0.25 % F.S.
BB480010
-500 to 500 mbar
0. 01 mbar ±0.25 % F.S.
BB480012
-0.95 to 2 bar
0.01 mbar
±0.25 % F.S.
BB480020
0 to 2 bar (ABS)
0.01 mbar
±0.25 % F.S.
BB480013
-0. 95 to 7 bar
0.1 mbar
±0.25 % F.S.
BB480014
-0. 95 to 20 bar
0.1 mbar
±0.25 % F.S.
BB480021
0 to 20 bar (ABS)
0.1 mbar
±0.25 % F.S.
BB480015
-0. 95 to 35 bar
1 mbar
±0.25 % F.S.
BB480016
0 to 70 bar
1 mbar
±0.25 % F.S.
BB480017
0 to 150 bar
1 mbar
±0.25 % F.S.
BB480018
0 to 350 bar
10 mbar
±0.25 % F.S.
BB480019
0 to 700 bar
10 mbar
±0.25 % F.S.
Overpressure: 125% F.S.
Port: ¼ BSPM (male)
Connection wire length: 2 meters
Environmental condition module (option) - Not available on IS models
Temperature:
Sensor type: Pt100
Range:
Accuracy:
Relative Humidity:
Sensor type:
Range: 0 to 95% RH
Accuracy: ±2% RH
Pressure:
Sensor type:
Range:
Accuracy: ±2 mbar
Math functions
Calculation functions: hold, max, min, offset, zero, average
In/Out data memory: 10 data with manual or automatic recall
Convert function: displays the electrical equivalent of the engineering unit
Scale factor: setting with zero and span programmable
Square root: in combination with scale factor
DataLogger
Sources: Voltage, Current, Temperature, Resistance
Sampling time: 1 sec
Memory: >1500 readings complete with date and time
General
Accuracy: the above accuracies are stated for 365 days and includes non-linearity,
hysteresis, and repeatability. The average temperature coefficient, inside the
temperature compensated range is ±0.002 of rdg/°C (w.t.r. +23°C / +73°F).
Compensation temperature range: 0 to 45°C (+32 to +113°F)
Calibration: self learning technique with automatic procedure
Common mode rejection: 140 dB at ac operation
Normal mode rejection: 60 dB at 50/60 Hz
Temperature stability:
for temperature exceeding the band +18°C to 28°C
Span: ±8 ppm/°C
Zero: ±0.2
µ
V/°C