3–58
3.4.3
CHECKING OF ONE SCALE POINT
3.4.3.1 Equipment
required
Scale point
NO
2
cylinder, the concentration of which is lower than 1 ppm (between 0.5 and 1 ppm) and is
connected to the span gas inlet of the analyzer.
Internal permeation bench equipped with a NO
2
tube (connected to the span gas inlet of the
analyzer). The concentration generated by the permeation bench is recorded on the test report.
Portable calibrator (VE3M type) equipped with a NO
2
tube and connected to the span gas inlet of
the analyzer.
NOTE :
If the analyzer is equipped with an internal permeation bench, the sample inlet must be
used to connect the reference gas cylinder or the portable calibrator.
3.4.3.2 Procedure
Scale point checking
By pressing the [Span] or [Sample] key, select the gas inlet where the span gas is connected and wait
for the measurement to stabilize. The measurement obtained must be compared to the concentration
generated by the equipment used, taking into account the accuracy of this equipment.
If the difference between the span gas value and the reading is higher than + 5 %, it is necessary to
carry out a span adjustment. Refer to section 0 «
MEASUREMENT
Instantaneous »
.
3.4.3.3
Use of automatic cycles
For programming cycles, refer to section 3.3.3.3 «
SPAN
Cycles »
.
Calibration cycle
The device for scale point generation is permanently linked to the span gas inlet of the analyzer. The
NO
2
titration has to be lower than the full scale of the range used in measurement. The recommended
timing for the check is 1200 seconds.
3.4.4
TWO-POINT CALIBRATION
3.4.4.1 Devices
Zero
air:
It is possible to use the zero air filter or, for better accuracy, a zero air generator with molecular
sieve or a synthetic air cylinder (mixture of 80 % N
2
and 20 % O
2
). These devices are linked to the
sample or span gas inlets of the analyzer.
Scale
point:
Cylinder of NO
2
, titrated at approximately 80 % of full scale of the measurement range used
(ac 1 %). For reasons of gas mixture stability, the use of a high-concentration (10 ppm or
more) cylinder is preferred. In this case, a dilution system (MGC101 type) has to be used to
decrease the gas concentrations to the usable range.
Summary of Contents for AS32M
Page 8: ...EnvironnementS A AS32M Duplication prohibited MAY 2013 0 8 Page intentionally left blank...
Page 26: ...EnvironnementS A AS32M Duplication prohibited AUGUST 2013 2 2 Page intentionally left blank...
Page 38: ...EnvironnementS A AS32M Duplication prohibited DECEMBER 2013 2 14 Page intentionally left blank...
Page 104: ...EnvironnementS A AS32M Duplication prohibited DECEMBER 2013 3 66 Page intentionally left blank...
Page 106: ...EnvironnementS A AS32M Duplication prohibited AUGUST 2013 4 2 Page intentionally left blank...
Page 111: ...Duplication prohibited AS32M EnvironnementS A AUGUST 2013 4 7 Page intentionally left blank...
Page 126: ...EnvironnementS A AS32M Duplication prohibited AUGUST 2013 5 2 Page intentionally left blank...
Page 133: ...Duplication prohibited AS32M EnvironnementS A DECEMBER 2013 5 9 Page intentionally left blank...
Page 140: ...EnvironnementS A AS32M Duplication prohibited DECEMBER 2013 5 16 Page intentionally left blank...
Page 142: ...EnvironnementS A AS32M Duplication prohibited AUGUST 2013 6 2 Page intentionally left blank...
Page 160: ...EnvironnementS A ESTEL Board Duplication prohibited JUNE 2009 18 Page intentionally left blank...