5–8
Alarm messages
Min./Max. values
Possible cause(s)
Possible action(s)
Permeation bench
temperature comprised
between 49°C and 51°C
–
The ambient temperature is out of
the operating limits (10°C – 51°c).
–
Permeation bench temperature
regulation is faulty.
–
Check supply of bench temperature
regulation board (+15 V).