Environnement
S.A
AS32M
Duplication prohibited
DECEMBER 2013
4–10
MAINTENANCE SHEET
MONITOR Serial No.:
OPERATION SHEET: 4.3.4
Scope: Check of optical and fluid parameters
PAGE: 1/3
Periodicity: 15 days
Date
Check of flow rate:
Connect a flow meter on the sample inlet of the AS32M.
Figure 4-4 – Flowmeter connection
Make sure the flow rate value indicated on the flow meter is 1 l/min (or 60 l/h).
Check of calibration coefficient:
In the «
SPAN
Calibration »
screen, make sure the calibration coefficient is between 0,8
and 1,3.
Check of chamber parameters
Switch the analyzer to zero air.
Compare the values given on the «
TESTS
Chamber parameters »
screen with the limits
listed in the table given next page.
Tool required
Flow meter, 100 liters/hour
Sample inlet
Flow meter
Summary of Contents for AS32M
Page 8: ...EnvironnementS A AS32M Duplication prohibited MAY 2013 0 8 Page intentionally left blank...
Page 26: ...EnvironnementS A AS32M Duplication prohibited AUGUST 2013 2 2 Page intentionally left blank...
Page 38: ...EnvironnementS A AS32M Duplication prohibited DECEMBER 2013 2 14 Page intentionally left blank...
Page 104: ...EnvironnementS A AS32M Duplication prohibited DECEMBER 2013 3 66 Page intentionally left blank...
Page 106: ...EnvironnementS A AS32M Duplication prohibited AUGUST 2013 4 2 Page intentionally left blank...
Page 111: ...Duplication prohibited AS32M EnvironnementS A AUGUST 2013 4 7 Page intentionally left blank...
Page 126: ...EnvironnementS A AS32M Duplication prohibited AUGUST 2013 5 2 Page intentionally left blank...
Page 133: ...Duplication prohibited AS32M EnvironnementS A DECEMBER 2013 5 9 Page intentionally left blank...
Page 140: ...EnvironnementS A AS32M Duplication prohibited DECEMBER 2013 5 16 Page intentionally left blank...
Page 142: ...EnvironnementS A AS32M Duplication prohibited AUGUST 2013 6 2 Page intentionally left blank...
Page 160: ...EnvironnementS A ESTEL Board Duplication prohibited JUNE 2009 18 Page intentionally left blank...