18
EN109-1665751-44
Figure 12
Comment:
the treadmill
can also be brought to a standstill by activating the:
- Safety belt with magnetic contact (see 9.2.10)
- Patient stop button (optional accessories; see 9.2.11)
- Autostop function (see 9.2.12)
9.2.6
Central controller (rotary knob)
Using the
central controller
[7]
,
•
you can select the desired menu functions (for example
Quick start
)
•
you can enter the values for the various parameters (for example
Weight
)
•
you can choose between several different options available in a screen (for example
Male/Female
).
Figure 13
9.2.7
Autospeed indicator lamp
The activation of the
Autospeed function [8]
means that the band speed will automatically adjust to the pace
of the person using the treadmill. If this person is unable to maintain a previously selected speed (and is
therefore at risk of ending up at the back of the running surface), the band speed will automatically decrease. If
the person is able to walk or run faster, the band speed will automatically increase. A green LED to the left of
the
AUTO button
indicates that the Autospeed sensor is activated (figure 14). For certain programmes and
exertion tests, Autospeed forms part of the standard test protocol. See 10.6 for a detailed description of the
Autospeed function.
Figure 14
9.2.8
Inclination buttons
The buttons for adjusting the incline (fig. 15) are on the right of the control panel. The selected incline can be
either positive (uphill) or negative (downhill). The range is expressed as a percentage, with max. +15%
(positive) and max. -10% (negative). The selected incline is displayed at the bottom right of the information
screen. For safety reasons, the treadmill will adjust gradually to the selected incline. A flashing arrow indicates
that the band has not yet reached the selected angle. An increasing angle of inclination is indicated by an
arrow pointing upwards. A decreasing angle of inclination is indicated by an arrow pointing downwards. During
the gradual adjustment process, the current angle of inclination will be shown.
Indicator lamp (LED)
STOP button [6]
- 1x press = stop/end
Central rotary controller [7]