Commissioning
Nanomass Gas Density
42
Hauser
Navigation
Menu "Measuring variables"
→
"Special units"
Parameter overview with brief description
Menu "Measuring variables"
→
"Special units"
10.5.2 Configuring reference density measurement
Settings for calculating the reference density can be made in the "Special Function" menu.
Navigation
Menu "Special function"
→
"Density function"
Parameter overview with brief description
Menu "Special function"
→
"Density function"
10.5.3 Configuring the concentration measurement
Information on concentration measurement
10.5.4 Performing density adjustment with medium
A density adjustment can be performed under process conditions to achieve optimum
accuracy. The density adjustment must be performed using the "Nanomass Communication"
operating tool.
Navigation
Function area "Device settings"
→
Menu "Calibrate density"
→
"Single point density calibration"
To access the "Device settings" function area, you must enter the access code
0074
in the
"Access code for technician mode" field and then press "OK" to confirm.
Parameter
Description
Selection/input
Factory setting
Text arb. density (4 char.)
Enter the text string for the arbitrary
density unit. The unit is then available
for selection in the "Density" parameter in
the "System unit" menu.
XXXX (max. 4 characters)
"___" (no text)
Factor arbitrary density
Enter the quantity factor for the arbitrary
density unit. The factor converts the
original measured value into the
arbitrary unit.
6-digit floating point number and
a 1-digit exponent
1.0
Parameter
Description
Selection/input
Factory setting
Density type
Select the density function for calculating
special density values or the percentage
of components in binary mixtures.
• Fixed
Off
Reference pressure
Enter the reference pressure for
calculating the reference density.
Decimal number with sign
14.7 psi a
Reference temperature
Enter the reference temperature for
calculating the reference density.
Decimal number with sign
20 °C