Technical data
Nanomass Gas Density
66
Hauser
Signal on alarm
Breakdown information is displayed as follows.
Current output
Local display
"Nanomass Communication" operating tool
Light emitting diodes (LED)
Galvanic isolation
The current outputs are galvanically isolated from the rest of the system.
16.5
Power supply
Pin assignment of
connector
(
21)
Supply voltage
Power consumption
Resolution
8
μ
A
Damping
Configurable: 0 to 120 s
Assignable measured variable
• Density
• Reference density
• Concentration
• Temperature
• Pressure
Failure mode
Choice of breakdown information (in accordance with NAMUR
Recommendation NE 43) if the tube does not oscillate correctly or a value
is outside the density range:
• Minimum value: 2 mA
• Maximum value: 22 mA (not Namur: 23 mA)
• Actual value
• Last valid value
Breakdown of power supply
Breakdown information during disconnection of power supply:
• <1 mA
Plain text display
Error message on operational display.
Status information
Error message with status information
Status information
Status indicated by two light emitting diodes (in accordance with NAMUR
Recommendation NE 43).
The light emitting diodes indicate the following information:
• Supply voltage active
• Data transmission active
• Device alarm/error has occurred
Power connection
DC 8 to 30 V, the power unit must be tested to ensure it meets safety
requirements (e.g. PELV, SELV)
USB port (optional)
DC 5 V
Maximum power consumption
400 mW