Commissioning
Nanomass Gas Density
46
Hauser
Menu "Special function"
→
"Density function"
Parameter
Description
Selection/input
Factory setting
Density type
Select the density function for calculating
special density values or the percentage
of components in binary gas mixtures.
• Flexible
Off
Concentration unit
Select the unit for calculating the
concentration.
• %Mass 3D
• %Volume 3D
• Other 3D
Other 3D
Concentration selection
Select the concentration specification.
The names that were defined in
the "Concentration names"
parameter for the concentration
specifications.
1st coeff
Concentration names
Enter the specific name for the
concentration specification.
Character string consisting of a
maximum of ten letters,
numbers, blank spaces and/or
special characters.
1st coeffs, 2nd coeffs, 3rd coeffs,
4th coeffs
A0
Enter the function coefficient A0.
6-digit floating point number and
a 1-digit exponent
0.0
A1
Enter the function coefficient A1.
6-digit floating point number and
a 1-digit exponent
0.0
A2
Enter the function coefficient A2.
6-digit floating point number and
a 1-digit exponent
0.0
A3
Enter the function coefficient A3.
6-digit floating point number and
a 1-digit exponent
0.0
A4
Enter the function coefficient A4.
6-digit floating point number and
a 1-digit exponent
0.0