17.11 Menu "Basic function"
→
"System parameter"
17.12 Menu "Special function"
→
"Density function"
Current value
Prerequisite
The "On" option is selected in the "Activate simulation" parameter.
Description
Use this function to enter a current value for the simulation.
User entry
2.00 to 23.00 mA
Factory setting
2.00 mA
Additional information
If simulation is enabled ("On" is selected), the message "Sim activ" appears
on the display of the local operation and the message "Current simulation
activ" appears in the header of the operating tool "Nanomass
Communication".
Time constant input
Description
Use this function to enter a response time of the local display to
fluctuations in the measured value.
User entry
Integer or decimal number 0 to 120 s
Factory setting
0 s
Additional information
The measure value is averaged over the time entered.
Density type
Prerequisite
The optional "Concentration Measurement" application package must be
used for the "Fixed" function.
Description
Use this function to select the density function for calculating special
density values or the percentage (concentration) of components in two-
phase media.
Options
• Off
• Fixed
• Flexible
Factory setting
Off
Additional information
If "Off" is selected,
the other functions are not displayed. If "Fixed" is
selected, a reference density of the medium can be output and if "Flexible"
is selected a concentration can be output.
Reference pressure
Prerequisite
This function is only available if the "Fixed" option
was selected in the
"Density type" function
.
Description
Use this function to enter a reference pressure for calculating the
reference density.
User entry
Decimal number with sign
Factory setting
14.7 psi a
Additional information
The unit depends on the unit selected in the "Pressure" parameter.