Appendix
Nanomass Gas Density
86
Hauser
17.13 Menu "Supervision"
→
"Assign error prompt for channel 2"
17.14 Menu "Device settings"
→
"Advanced con
fi
guration"
Concentration offset
Prerequisite
This function is only available if the "Flexible" option was selected in the
"Density type" function.
Description
Use this function to enter an offset value for the calculated concentration.
User entry
Decimal number with sign -5.000 to 5.000
Factory setting
0.0
Additional information
The offset is added or subtracted, depending on the sign.
Formula coefficients
Prerequisite
This function is only available if the "Flexible" option was selected in the
"Density type" function.
Description
Use this function to enter the coefficients A0 to A4.
User entry
Floating point number with sign
Factory setting
0.0
Tube not oscillating
Description
Use this function to select whether an error message should be displayed
at current output 2 if the measuring channel is not oscillating or is
oscillating in the wrong frequency range.
Options
• On
• Off
Factory setting
Off
Density range
Description
Use this function to select whether an error message should be displayed
at current output 2 if the density value is outside the measuring range.
Options
• On
• Off
Factory setting
Off
Restart device
Description
Use this function to select whether the measuring device should be
restarted.
Options
Local operation:
• Yes
• No
Factory setting
Local operation: No
Additional information
Local operation: If "Yes" is selected, the measuring device is restarted.
Restart the device via operating tool "Nanomass Communication":
1. Click the "Perform restart" button.
A window appears.
2. Click the "Continue" button to restart the measuring device.