Nanomass Gas Density
Installation
Hauser
19
2. Connect the filter to the pipe (
15) and secure with a Swagelok pipe fitting (
18).
3. Mount the measuring device on a wall or ceiling using drill holes (
16).
4. Connect the measuring device to the pipe and secure with a Swagelok pipe fitting (
18).
6.3
Post-installation check
Is the device undamaged (visual inspection)?
If assemblies are installed, are they undamaged (visual inspection)?
Does the measuring device conform to the measuring point specifications?
For example:
• Medium temperature range
• Process pressure (refer to the section on "Pressure-temperature ratings" in the "Technical
Information" document )
• Ambient temperature range
• Measuring range
Are the measuring point identification and labeling correct (visual inspection)?
Is the measuring device adequately protected from precipitation and direct sunlight?
Are the securing screws tightened securely?
Is the pressure sensor connected?