Commissioning
Nanomass Gas Density
36
Hauser
10
Commissioning
10.1
Installation and function check
Before commissioning the device, make sure that the post-installation and post-connection
checks have been performed.
• "Post-installation check" checklist
19.
• "Post-connection check" checklist
10.2
Switching on the measuring device
After a successful function check, switch on the measuring device. If power is supplied via
the USB port, the measuring device starts automatically as soon as the USB cable is
connected.
After a successful startup, the local display switches automatically from the startup display
to the operational display.
If nothing appears on the local display or an error message is displayed, refer to the
section on "Diagnostics and troubleshooting"
54.
10.3
Setting the operating language
The operating language of the measuring device is English. Other language options are not
available.
10.4
Configuring the measuring device
An overview of all the device parameters and a description of the parameters is provided
in the appendix
75.
10.4.1 Defining the tag name
The tag name must be defined via the "Nanomass Communication" operating tool.
Navigation
Function area "Device settings"
→
"Device information"
→
"Tag name"
To access the "Device settings" function area, you must enter the access code
0074
in the
"Access code for technician mode" window and then press "OK" to confirm.
Defining the tag name
1. Call up the "Device settings" function area.
2. Enter the new tag name under "Tag name" in the "Device information" area.
3. Click "Send" to confirm the name.
The message "Tag name changed" appears.
The name is saved in the device.
4. Click "OK" to confirm the message.
10.4.2 Setting the serial port delay time
The delay time for sending and receiving data via the serial port must be defined via the
"Nanomass Communication" operating tool.