10.4.6 Configuring the local display
Various parameters are used to configure the local display.
Navigation
1. Menu "User interface"
→
"Assign"
→
"Top line display"
2. Menu "User interface"
→
"Assign"
→
"Top line display multiplex"
3. Menu "User interface"
→
"Assign"
→
"Bottom line display"
4. Menu "User interface"
→
"Assign"
→
"Bottom line display multiplex"
Parameter overview with brief description
Menu "User interface"
→
"Assign"
→
"Top line display"
Menu "User interface"
→
"Assign"
→
"Top line display multiplex"
Failsafe mode channel 2
Specify the output behavior of current
output 2 in the event of a device alarm.
Note
In the event of a breakdown of the power
supply the value of the current output is
<1 mA. The failure mode in accordance
with NAMUR Recommendation NE 43 is
ensured.
• Min current
• Max current
• Hold value
• Actual value
Min current
Parameter
Description
Selection/input
Factory setting
Parameter
Description
Selection/input
Factory setting
Tube not oscillating
Select whether an error message should
be displayed at current output 2 if the
measuring channel is not oscillating or is
oscillating in the wrong frequency range.
• On
• Off
Off
Density range
Select whether an error message should
be displayed at current output 2 if the
density value is outside the permitted
range.
• On
• Off
Off
Parameter
Description
Selection/input
Factory setting
Top line display
Select the measured value that is shown
on the top line of the local display.
• Off
• Density
• Concentration (optional)
• Ref. density
• Temperature
• Pressure
Density
Parameter
Description
Selection/input
Factory setting
Top line display multiplex
Select the measured value that is shown
on the top line of the local display when
the display changes.
• Off
• Density
• Concentration (optional)
• Ref. density
• Temperature
• Pressure
Off