Procedure
1. Make sure that a pressure regulator is connected to the nitrogen gas bottle.
2. Connect a hose from the nitrogen gas bottle through a T-piece to the sample supply
port on the rear of the analyzer. Connect an excess flow line to the unused port on
the T-piece and route the excess flow line to a suitable extractor.
3. On the display controller of the analyzer, browse to the
Main menu
as described in
Figure 6-2: Main Menu
4. Using
and
, select
GAS SERVICE
.
5. Press
.
The
Select gas
screen (
Figure 6-3: Select gas Screen
6. Use
and
to move the cursor until the gas that you wish to zero calibrate is
selected. Press
.
Reference Manual
Gas calibration
00809-0100-5800
September 2019
Rosemount CT5800
83