Figure 6-1: Main menu
Use the
Main menu
to access the software routines and screens that are used for
calibration and maintenance. Five options are presented (
information about the options on the
Main menu
, see
.
On the
Main menu
, use
and
to scroll between the menu options.
When the option you want is highlighted (
GAS SERVICE
is the example shown in
), press
to go to the first screen of that software routine.
6.3
Calibration
Calibration flows the known cylinder gas through the analyzer and then adjusts the
readout until the measurement matches the cylinder. The analyzer then applies this
adjustment to all measurements until the next calibration. There are two circumstances in
which you may want to calibrate.
1. After you verify gas concentrations (see
) and find that
the analyzer readings are outside of tolerance
2. At regular intervals, such as once a day or once a shift
6.3.1
Zero calibration
Zero calibration is calibrating the analyzer so that when no sample gas is flowing through
it, the gas concentrations measured by the analyzer are zero. To zero calibrate, calibrate
the analyzer measurements against a known sample gas using the following procedure:
Prerequisites
Run the analyzer at a stable temperature for at least thirty minutes prior to commencing
this procedure.
Use nitrogen gas of "instrument gas" purity as the zero calibration gas.
Gas calibration
Reference Manual
September 2019
00809-0100-5800
82
Emerson.com/Rosemount
Summary of Contents for Rosemount CT5800
Page 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Page 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Page 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Page 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Page 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Page 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Page 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...