1.9
Detailed system specifications
gives the physical characteristics of the analyzer.
diagrams of the sensor and mounting points.
gives the general characteristics of
the analyzer.
Table 1-2: Physical Characteristics
Rosemount CT5800
Value
Comment
External dimensions (closed)
694.5 x 292 x 515 mm
27.34 x 11.5 x 20.28 in.
Length x Width x Height
Nominal dimensions
External dimensions (open)
27.34 x 11.5 x 41.24 in.
694.5 x 292 x 1047.5
mm
Nominal dimensions, front panel at lowest point
To open the enclosure, the minimum height required is
1030 mm (40.55 in.)
Front panel swept radius
21.46 in.
545 mm
Allow sufficient space for the front panel to swing through
an arc of 180 ° (see
).
Weight
80 kg
176.37 lb.
Approximate weight
The analyzer must be wall mounted. Do not mount it on
stud or partitioned walls.
Table 1-3: General Characteristics
Rosemount CT5800
Value
Units
Comment
Voltage
220/240 ± 10%
110/120 ± 10%
Vac
50 Hz ± 5% (factory set)
60 Hz ± 5% (factory set)
Peak power consumption
1100
W
Max consumption per gas analyzer
Continuous steady-state
power consumption
800
W
Electrical compartment
enclosure
-
-
Aluminum
Optical compartment
enclosure
-
-
Aluminum
Measurement technique
-
-
Mid IR absorption spectroscopy
Mid IR source
N/A
N/A
Quantum Cascade Laser
Near IR source
N/A
N/A
Interband Cascade Laser Tunable diode laser
Laser classification
Class 1
BS EN 60825-1: 2007 safety of laser products.
Equipment classification and requirements
(identical to IEC 60825-1 2007)
Inlet gas port connector
¼
6
in.
mm
Swagelok
®
type, factory-configured, specify on
order
Outlet (exhaust) gas port
connector
¼
6
in.
mm
Swagelok type, factory-configured, specify on order
Measurement result signals
4 to 20
or Modbus over TCP /
IP
mA
4 or 8 channel outputs, specify on order
Plan
Reference Manual
September 2019
00809-0100-5800
28
Emerson.com/Rosemount
Summary of Contents for Rosemount CT5800
Page 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Page 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Page 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Page 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Page 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Page 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Page 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...